{"recall":{"id":"fda-device-99371","source":"FDA","sourceName":"U.S. Food and Drug Administration (openFDA device enforcement)","cat":"medicines","kids":false,"kind":"device","title":"Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI)","firm":"Fort Defiance Industries, LLC","date":"2026-07-08","reported":"2026-08-26","status":"Ongoing","classification":"Class II","products":[{"name":"Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131","code":"Lot Code: All serial numbers/ GTIN: (01) 00862992000327","quantity":"","number":"Z-2983-2026","classification":"Class II"}],"reason":"The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.","hazard":"FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).","affected":"Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.","nationwide":true,"states":[],"units":"","remedy":"","whatToDo":"Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.","contact":"","url":"https://www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99371","apiUrl":"https://api.fda.gov/device/enforcement.json?search=report_date:[20260705+TO+20991231]&sort=report_date:desc&limit=1000","rawHash":"6ad9a15bcdede685","firstSeen":"2026-11-02T13:57:14.206Z","updatedAt":"2026-11-02T13:57:14.206Z","history":[]},"official":{"fetchedAt":"2026-11-02T13:57:14.206Z","apiUrl":"https://api.fda.gov/device/enforcement.json?search=report_date:[20260705+TO+20991231]&sort=report_date:desc&limit=1000","officialUrl":"https://www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99371","record":[{"status":"Ongoing","city":"Loudon","state":"TN","country":"United States","classification":"Class II","openfda":{},"product_type":"Devices","event_id":"99371","recalling_firm":"Fort Defiance Industries, LLC","address_1":"2411 Maremont Pkwy","address_2":"","postal_code":"37774-1066","voluntary_mandated":"Voluntary: Firm initiated","initial_firm_notification":"","distribution_pattern":"Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.","recall_number":"Z-2983-2026","product_description":"Fort Defiance Industries (FDI),  Automated Field Steam Sterilizer  REF: P2131","product_quantity":"","reason_for_recall":"The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.","recall_initiation_date":"20260708","center_classification_date":"20260818","report_date":"20260826","code_info":"Lot Code: All serial numbers/  GTIN: (01) 00862992000327","more_code_info":""}]}}