OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Steris Corporation recalls 6 products

Recall date · record updated Oct 14, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Steris Corporation
Reason
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99704…

Products (6)

  1. VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

    Codes/lots: UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027Quantity: 266Recall no. Z-0070-2027Class II
  2. VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

    Codes/lots: UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027Quantity: 409Recall no. Z-0071-2027Class II
  3. VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

    Codes/lots: UDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027Quantity: 1185Recall no. Z-0072-2027Class II
  4. VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

    Codes/lots: UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027Quantity: 2205Recall no. Z-0073-2027Class II
  5. VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

    Codes/lots: UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027Quantity: 217Recall no. Z-0074-2027Class II
  6. VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

    Codes/lots: UDI:10724995023864 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027Quantity: 233Recall no. Z-0075-2027Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls