Recalls › Medicines & devices · Kids & Babies
Class IIOngoing
Medline Industries, LP recalls 3 products
Recall date · record updated Sep 2, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Medline Industries, LP
- Reason
- Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide distribution - US Nationwide and the country of Panama.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99403…
Products (3)
Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.
Codes/lots: All lot codesQuantity: 20 unitsRecall no. Z-2991-2026Class IIMedline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).
Codes/lots: 1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344Quantity: 145525 unitsRecall no. Z-2992-2026Class IIMedline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).
Codes/lots: 1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot nQuantity: 271406 unitsRecall no. Z-2993-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass II
American Health Packaging recalls Metoclopramide Tablets
Presence of Foreign Tablets/Capsules
· FDA · Nationwide
Medicines & devicesClass II
Zydus Pharmaceuticals (USA) Inc recalls Hydrocortisone Sodium Succinate for Injection
Failed Impurities/Degradation Specifications.
· FDA · Nationwide
Medicines & devicesClass II
Teva Pharmaceuticals USA, Inc recalls Metoclopramide Tablets
Presence of Foreign Tablets/Capsules
· FDA · Nationwide