OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Cadwell Industries Inc recalls Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 0084006710

Recall date · record updated Oct 21, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Cadwell Industries Inc
Reason
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99774…

Products

  1. Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electrical stimulator, a TCS stimulator, a visual stimulator and an auditory stimulator. The low current stimulator has one (1) constant current output pair with a current range of 0.01-20 mA. The TCS stimulator has nine (9) independent constant current/constant voltage outputs, with current range of 50-1500 mA and voltage range of 50-1000 V. Component: Is the cortical component of the IOMAX system

    Codes/lots: Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Lot Code: These devices are identified by the serial numbers listed below: 19029607AA0326007 19029607AA0226013 19029607AA0426028 19029607AA0226010 19029607AA0226007 19029607AA0326026 19029607AA0226015 19029607AA0326020 19029607AA0326018 19029607AA0326021 19029607AA0326022 19029607AA0226009 19029607AA0526022 19029607AA0326017 19029607AA0326027 19029607AA0326016 19029607AA0226016 19029607AA0326028 19029607AA0326024 19029607AA0326025 19029607AA0326023 19029607AA0326031 19029607AA0326006Quantity: 23Recall no. Z-0141-2027Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls