OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Medacta Usa Inc recalls Short Monobloc Stem Broaches

Recall date · record updated Jul 8, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Medacta Usa Inc
Reason
Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=98944…

Products

  1. Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System

    Codes/lots: REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095Quantity: 50Recall no. Z-2618-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.