OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Alphatec Spine, Inc. recalls SafeOp Patient Interface

Recall date · record updated Jul 29, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Alphatec Spine, Inc.
Reason
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring system reboot; may result in increased operative time and surgical plan change.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
US Nationwide distribution in the states of TX, MO, IN.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=98967…

Products

  1. SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System

    Codes/lots: UDI-DI: 00190376517506. System Set ID/Patient Interface Serial Number: 2007716/ DG106225, 2005498/ DG106220, 2005490/ DG106230Quantity: 3Recall no. Z-2793-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API