OVERLOOKED

Recalls › Medicines & devices

Class IOngoingUpdated Nov 2, 2026

Insulet Corporation recalls 2 products

Recall date · record updated Jul 8, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Insulet Corporation
Reason
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Hazard
FDA Class I: a reasonable probability that use or exposure will cause serious adverse health consequences or death (FDA definition).
Who’s affected
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99034…

Products (2)

  1. Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).

    Codes/lots: 1. Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U1011Quantity: 4,936,056 unitsRecall no. Z-2603-2026Class I
  2. Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).

    Codes/lots: 1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PQuantity: 1,790,784 unitsRecall no. Z-2604-2026Class I

Update history

  1. First listed on OVERLOOKED from the official record.
  2. Product details, codes or quantities changed; Units: 4,936,056 units; 1,790,784 units; 289,046 units → 4,936,056 units; 1,790,784 units

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls