Class IIOngoing
Accuray Incorporated recalls Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer
Recall date · record updated Jul 8, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Accuray Incorporated
- Reason
- In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99083…
Products
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Codes/lots: ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table.Quantity: 292 unitsRecall no. Z-2644-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide