OVERLOOKED

Recalls › Medicines & devices

Class IOngoing

Medline Industries, LP recalls Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no

Recall date · record updated Jul 8, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Medline Industries, LP
Reason
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Hazard
FDA Class I: a reasonable probability that use or exposure will cause serious adverse health consequences or death (FDA definition).
Who’s affected
US Nationwide distribution in the state of TX.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99122…

Products

  1. Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.

    Codes/lots: Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637.Quantity: 3Recall no. Z-2615-2026Class I

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls