OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Integra LifeSciences Corp. (NeuroSciences) recalls OmniTract Surgical Retractor System

Recall date · record updated Jul 15, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Integra LifeSciences Corp. (NeuroSciences)
Reason
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Korea (the Republic of), Kuwait, Lithuania, Malaysia, Malta, Mauritius, Mexico, Nepal, Netherlands (the), New Zealand, Norway, Pakistan, Palestine, State of, Philippines (the), Poland, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, R¿union, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan (Province of China), Thailand, Trinidad and Tobago, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland, (the), Viet Nam.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99193…

Products

  1. OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

    Codes/lots: Product Code 3032, UDI 10381780268468; Product Code 3055, UDI 10381780268505; Product Code 3069, UDI 10381780268512; Product Code 3076, UDI 10381780268529; Product Code 3086, UDI 10381780268536; Product Code 3092, UDI 10381780268550; Product Code 3093, UDI 10381780268567; Product Code 3096, UDI 10381780268581; Product Code 3098, UDI 10381780268598; Product Code 3099, UDI 10381780268604; Product Code 3133, UDI 10381780268680; Product Code 3137, UDI 10381780268697; Product Code 3142, UDI 10381780268710; Product Code 3146, UDI 10381780268734; Product Code 3148, UDI 10381780268758; Product Code 31Quantity: 633 unitsRecall no. Z-2714-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls