OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Medline Industries, LP recalls 8 products

Recall date · record updated Jul 8, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Medline Industries, LP
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99209…

Products (8)

  1. Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B

    Codes/lots: UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720Quantity: 9,272 kitsRecall no. Z-2606-2026Class II
  2. Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N

    Codes/lots: DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783Quantity: 9,272 kitsRecall no. Z-2607-2026Class II
  3. Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407

    Codes/lots: UDI-DI 10198459704543 Lot 26DBB599Quantity: 9,272 kitsRecall no. Z-2608-2026Class II
  4. Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B

    Codes/lots: DYNJ905855 UDI-DI 10193489546606 Lots 21ABU297 21BBB235 21DBV520 21FBG162 21GBC672 21GBI192 21IBT387 21VBB580 22DBK812 22FBR093 22FBS035 22HBI628 22IBV291 22JBX729 22NBE430 23ABD954 23BBI465 23CBO951 23CBX474 23GBF721 DYNJ905857 UDI-DI 10193489545678 Lots 21ABG510 21GBK940 DYNJ905936B UDI-DI 10193489422818 Lots 21BBF766 21DBP452 21FBF248 21LBJ459 22ABV319 22DBS751 22FBU868Quantity: 9,272 kitsRecall no. Z-2609-2026Class II
  5. Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G

    Codes/lots: DYNJ905503D UDI-DI 10193489430295 lots 21DBG611 21EBI605 21FBN751 DYNJ905503F UDI-DI 10193489967272 lots 21HBL060 21IBW859 22DBC340 22DBS437 22FBR267 22GBR286 22HBS560 22IBA577 22JBI883 22KBD587 22LBF425 23BBN903 23CBL450 23CBW776 23FBC209 23IBU343 24DBQ181 24FBT062 24GBR740 24HBI428 24JBP984 DYNJ905503G UDI-DI 10198459200397 lots 25ABJ069 25ABL749 25CBB762 25DBG799 25EBM994 25GBW226 25LBA950 25LBT019 26ABH519 26ABT315 26BBQ338 26DBJ110Quantity: 9,272 kitsRecall no. Z-2610-2026Class II
  6. Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M

    Codes/lots: DYNJ909090B UDI-DI 10195327652432 Lots 24EBL051 24GBG271 CDS982523M UDI-DI 10195327560591 Lots 24ALA227 24BLA369 24CLA108 24CLA518 24DLA082 24ELB053 24ILA112Quantity: 9,272 kitsRecall no. Z-2611-2026Class II
  7. Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A

    Codes/lots: UDI-DI 10195327033620 Lots 21KBN891 21KBO410 22ABA676 22ABY937 22BBN909 22BBT447 23ABO303Quantity: 9,272 kitsRecall no. Z-2612-2026Class II
  8. Medline Convenience kits containing recalled Swan-Ganz Catheters CABG A CDS CDS982269K CDS982269L CDS982269M CDS982269N CARDIAC MAJOR DYNJ901328D DYNJ901328F HEART A DYNJ903465F DYNJ903465G HEART TRAY-VASCULAR DYNJ907857 DYNJ907857A DYNJ907857B KIT UNIVERSITY CABG PK DYNJ905857A OPEN HEART CDS981161AA CDS981161AB DYNJ905437A DYNJ909090 DYNJ909090C DYNJ9855016R DYNJ9855016S OPEN HEART ACCESSORY CDS CDS984289J OPEN HEART ANESTHESIA PACK CDS983849D CDS983849F CDS983849G OPEN HEART B CDS DYNJ900377P OPEN HEART BASIC B DYNJ904311C OPEN HEART CDS CDS983513F OPEN HEART CDS QVH CDS982523K CDS982523L CDS982523N OPEN HEART LINE SET UP DYNJ908179 DYNJ908179A OPEN HEART SUPPLEMENT DYNJ909116 DYNJ909116A DYNJ909116C DYNJ909116D SILVER CROSS OPEN HEART DYNJ906102J DYNJ906102K DYNJ906102L SM OPEN HEART PACK A&B DYNJ904749G DYNJ904749J VPH OPEN HEART DYNJ902664J DYNJ902664K DYNJ902664L

    Codes/lots: CDS982269K UDI-DI 10193489906844 Lots 21HBN691 21JBQ518 21LBD444 CDS982269L UDI-DI 10195327018016 Lots 21LBP501 22ABR514 22BBE936 22FBV989 CDS982269M UDI-DI 10195327192808 Lots 22JBE028 22LBF124 23ABP909 23CBO580 CDS982269N UDI-DI 10195327352998 Lots 23GBC969 23HBS959 23KBW067 23LBF510 24CBO691 24DBR250 24EBU619 24GBF288 25ABM339 25CBJ666 25HBH657 25IBM477 26BBG451 DYNJ901328D UDI-DI 10193489854923 Lots 21CBQ397 21FBJ935 21HBR220 21JBV897 21JBV898 22ABG611 DYNJ901328F UDI-DI 10195327044572 Lots 22CBX749 22FBX103 23ABC195 23EBG860 23EBL637 23EBS968 23FBM688 23GBS043 24EBL355 24FBJ301 24IBS567 2Quantity: 9,272 kitsRecall no. Z-2613-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls