OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Canon Medical System, USA, INC. recalls Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used

Recall date · record updated Sep 2, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Canon Medical System, USA, INC.
Reason
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99222…

Products

  1. Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

    Codes/lots: Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: PendiQuantity: 10 systemsRecall no. Z-3014-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.