OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

3B Medical, Inc. recalls Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Model/Catalog Number: RII1001 (Small)

Recall date · record updated Jul 22, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
3B Medical, Inc.
Reason
Packaging and labeling discrepancy involving product size identification and associated Unique Device Identifier (UDI/GTIN).
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
US Nationwide distribution.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99247…

Products

  1. Brand Name: Rio" II Product Name: Rio" II Nasal Pillows Interface Model/Catalog Number: RII1001 (Small), RII1002 (Medium), RII1003 (Large) Software Version: N/A Product Description: Nasal pillows interface intended for use with positive airway pressure therapy devices. The device delivers airflow from a PAP device to the patient's nose and is secured using adjustable headgear. Component: No

    Codes/lots: Lot Code: Affected Catalog Numbers: RII1001, RII1002, RII1003 Affected Lots RII1001: 220121 220121 220191 RII1002: 220121 220121 220191 RII1003: 220121 220121 220191Quantity: 375Recall no. Z-2740-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls