Boston Scientific Corporation recalls 4 products

Recall date · record updated Jul 29, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Boston Scientific Corporation
Reason
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Global Distribution. US Nationwide.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99252…

Products (4)

  1. FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01

    Codes/lots: UDI-DI: 00810087181062; Lots: CL12910, CL12978, CL15153Quantity: 400 unitsRecall no. Z-2754-2026Class II
  2. FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062

    Codes/lots: UDI-DI: 00810087180072; Lots: CL12697, CL12698, CL12701, CL12702, CL12824, CL12832, CL12833, CL12861, CL12888, CL12889, CL12912, CL12913, CL12916, CL12918, CL12968, CL12975, CL12977, CL13003, CL13016, CL13019, CL13045, CL13047, CL13059, CL13078, CL13079, CL13101, CL13102, CL13120, CL13121, CL13140, CL13141, CL13158, CL13159, CL13195, CL13211, CL13218, CL13232, CL13233, CL13259, CL13290, CL13339, CL13367, CL13400, CL13421, CL13516, CL13518, CL13572, CL13577, CL13615, CL13645, CL13669, CL13707, CL13720, CL13733, CL13821, CL13834, CL13857, CL13926, CL13941, CL13976, CL13989, CL14026, CL14052, CL1Quantity: 74,760 unitsRecall no. Z-2755-2026Class II
  3. FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

    Codes/lots: UDI-DI: 00191506043131; Lots: CL12695, CL12696, CL12699, CL12700, CL12703, CL12808, CL12816, CL12817, CL12823, CL12834, CL12835, CL12836, CL12853, CL12860, CL12882, CL12883, CL12884, CL12891, CL12907, CL12908, CL12909, CL12919, CL12949, CL12950, CL12969, CL12976, CL12979, CL12980, CL13005, CL13010, CL13011, CL13012, CL13018, CL13032, CL13033, CL13034, CL13037, CL13056, CL13057, CL13058, CL13074, CL13075, CL13076, CL13077, CL13097, CL13098, CL13099, CL13100, CL13104, CL13115, CL13116, CL13117, CL13118, CL13119, CL13135, CL13136, CL13137, CL13138, CL13139, CL13156, CL13157, CL13191, CL13192, CL1Quantity: 508,151 unitsRecall no. Z-2756-2026Class II
  4. FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

    Codes/lots: UDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142, CL13160, CL13196, CL13212, CL13219, CL13231, CL13260, CL13291, CL13340, CL13375, CL13401, CL13422, CL13468, CL13519, CL13573, CL13578, CL13614, CL13646, CL13670, CL13671, CL13708, CL13719, CL13722, CL13732, CL13760, CL13801, CL13802, CL13820, CL13833, CL13856, CL13910, CL13911, CL13925, CL13940, CL13971, CL13972, CL13988, CL14025, CL14039, CL14050, CL14051, CL14076, CL14077, CL14119, CL14120, CL14135, CL14137, CL14171, CL14187, CL14226, CL14249, CL14286, CL14287, CL14345, CL1Quantity: 193,000 unitsRecall no. Z-2757-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.