Class IIOngoing
Pride Mobility Products Corp. recalls 13 products
Recall date · record updated Aug 12, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Pride Mobility Products Corp.
- Reason
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99286…
Products (13)
Brand Name: Quantum Rehab Product Name: Quantum 4Front 2 Model/Catalog Number: Quantum 4Front 2 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Quantum 4Front 2 0060650920019*6 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2841-2026Class IIBrand Name: Quantum Rehab Product Name: Quantum Outback Model/Catalog Number: Quantum Outback Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Quantum Outback 0060650920024*0 K112815 D143578 41875 ITIQuantity: 1102Recall no. Z-2842-2026Class IIBrand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge 3 Stretto 0060650920018*9 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2843-2026Class IIBrand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: R-TRAK 0060650920022*6 K011993 D049085 41875 ITIQuantity: 1102Recall no. Z-2844-2026Class IIBrand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge HD 0060650920005*9 K042612 D049086 41875 ITIQuantity: 1102Recall no. Z-2845-2026Class IIBrand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge 2.0 0060650920004*2 K042612 D049086 41875 ITIQuantity: 1102Recall no. Z-2846-2026Class IIBrand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Quantum 1450 0060650920009*7 K112815 D143578 41875 ITIQuantity: 1102Recall no. Z-2847-2026Class IIBrand Name: Quantum Rehab Product Name: Q6 Edge 2.0 (ilevel) Model/Catalog Number: Q6 Edge 2.0 (ilevel) Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge 2.0 (ilevel) 0060650920004*2 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2848-2026Class IIBrand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge 3 0060650920015*8 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2849-2026Class IIBrand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Q6 Edge 4 0060650920028*8 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2850-2026Class IIBrand Name: Quantum Rehab Product Name: Quantum 4Front 2HD Model/Catalog Number: Quantum 4Front 2HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Quantum 4Front 2HD 0060650920020*2 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2851-2026Class IIBrand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: 4Front 0060650920003*5 K143383 D246699 41875 ITIQuantity: 1102Recall no. Z-2852-2026Class IIBrand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Codes/lots: Lot Code: Quantum Rival 0060650920013*4 K011993 D049085 41875 ITIQuantity: 1102Recall no. Z-2853-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide