OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Ferno-Washington Inc recalls Power F2 Powered Fastening System

Recall date · record updated Jul 29, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Ferno-Washington Inc
Reason
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99291…

Products

  1. Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

    Codes/lots: UDI-DI 00190790006570 Serial Numbers: 25C227687 25C228204 26C232382 26C233091 25C227688 25C228205 26C232383 26C233092 25C227689 26C231797 26C232384 26C233093 25C227690 26C231798 26C232709 26C233094 25C227691 26C231799 26C232710 26C233095 25C227692 26C231800 26C232711 26C233096 25C227693 26C231801 26C232712 26C233097 25C227694 26C232002 26C232713 26C233098 25C227695 26C232004 26C232714 26C233099 25C227696 26C232005 26C232715 26C233100 25C227697 26C232006 26C232716 26C233111 25C227698 26C232007 26C232717 26C233112 25C227699 26C232008 26C232718 26C233113 25C227700 26C232009 26C233032 26C233114 25Quantity: 278Recall no. Z-2760-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls