Class IIOngoing
Copan WASP recalls Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2
Recall date · record updated Aug 12, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Copan WASP
- Reason
- Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- US Nationwide distribution in the state of Arizona.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99302…
Products
Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
Codes/lots: Lot Code: WebApp version 2.33.1, 2.33.3 and 2.34.2Quantity: 1Recall no. Z-2854-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide