Class IIOngoing
MUNEVO GMBH recalls munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/
Recall date · record updated Aug 12, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- MUNEVO GMBH
- Reason
- A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99328…
Products
munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
Codes/lots: Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699Quantity: 19Recall no. Z-2856-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide