OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Boston Scientific Corporation recalls Rapid Refill Continuous Injection System

Recall date · record updated Jul 22, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Boston Scientific Corporation
Reason
Devices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
US Nationwide distribution. OUS consignee list pending.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99336…

Products

  1. Rapid Refill Continuous Injection System, UPN M00566001

    Codes/lots: UDI-DI (GTIN) 08714729285151 Lots 33786611, 33825949, 33826026, 33949916, 33968571, 33968575, 34079981, 34153104, 34153413, 34153415, 34203075, 34203077, 34153411, 34203079, 34203115, 34386605, 34386609, 34386613, 34429656, 34429658, 34595249, 34693559, 34616613, 34616615, 34712433, 34763427, 34763967, 34805723, 34805725, 34805729, 35006995, 35006999, 35056793, 35208946, 35186845, 35208950, 35397314, 35208952, 35305436, 35498030, 35448316, 35498028, 35551523, 35626939, 35574463, 35688118, 35688460, 35828714,35831703, 35848483, 35877768, 35995282, 36049415, 36159194, 36219882, 36294100, 3630361Quantity: 153,921 (US); 119,274 (OUS)Recall no. Z-2725-2026Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API

More Medicines & devices recalls