OVERLOOKED

Recalls › Medicines & devices

Class IIIOngoing

Siemens Healthcare Diagnostics, Inc. recalls Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No

Recall date · record updated Aug 12, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Siemens Healthcare Diagnostics, Inc.
Reason
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Hazard
FDA Class III: not likely to cause adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99381…

Products

  1. Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

    Codes/lots: UDI-DI 00842768014321 Lot U2Quantity: 494 US; 114 OUSRecall no. Z-2881-2026Class III

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API