Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin REF: 11537225

Recall date · record updated Aug 12, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Siemens Healthcare Diagnostics, Inc.
Reason
Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99389…

Products

  1. Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

    Codes/lots: All Lots within expiry/UDI: 00630414611099Quantity: 7018 unitsRecall no. Z-2890-2026Class II