Class IIOngoing
Heidelberg Engineering GmbH recalls Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610
Recall date · record updated Sep 30, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Heidelberg Engineering GmbH
- Reason
- Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99595…
Products
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye
Codes/lots: UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611Quantity: 87Recall no. Z-3215-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide