OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Percussionaire Corporation recalls 7 products

Recall date · record updated Oct 7, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Percussionaire Corporation
Reason
Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99617…

Products (7)

  1. Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual

    Codes/lots: Product Number/UDI-DI: S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev DQuantity: 61,858Recall no. Z-3319-2026Class II
  2. In-Line Valve, Product Number: P5-TEE

    Codes/lots: UDI: 00849436000723Quantity: 98,434Recall no. Z-3320-2026Class II
  3. Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual

    Codes/lots: Product Number/UDI-DI: S00002-C/00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648; P5/00849436000501; Instruction Manual: P20045 Rev BQuantity: 229,105Recall no. Z-3321-2026Class II
  4. Intrapulmonary Percussive Ventilation, IPV-2C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual

    Codes/lots: Product Number/UDI-DI: S00002-C/00849436000020, 00849436000563, 00849436000570, 00849436000143, 00849436000648; P5/00849436000501; Instruction Manual: P20046 Rev AQuantity: 229,430Recall no. Z-3322-2026Class II
  5. Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual

    Codes/lots: Product Number/UDI-DI: S00010/00849436000389, P5/00849436000501, Instruction Manual: P20047 Rev EQuantity: 228,277Recall no. Z-3323-2026Class II
  6. Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual

    Codes/lots: Product Number/UDI-DI: S00055/00849436000242, P5/00849436000501, Instruction Manual: P20106 Rev BQuantity: 228,596Recall no. Z-3324-2026Class II
  7. Intrapulmonary Percussive Ventilation, IPV HC, Product Number: S00009; Phasitron, P5-HC

    Codes/lots: Product/Catalog Number/UDI-DI: S00009/00849436000358, P5-HC/00849436000525Quantity: 419Recall no. Z-3325-2026Class II