Class IIOngoing
Medical Action Industries, Inc. 306 recalls Blood Culture Collection Venipuncture Kit 20/Cs
Recall date · record updated Sep 23, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Medical Action Industries, Inc. 306
- Reason
- Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- US Nationwide distribution in the states of OH, IN, NC.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99628…
Products
Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood Culture Collection Kit, REF: 81834C. All containing BD ChloraPrep Clear - 1 mL Applicators
Codes/lots: 77451/20809160195069/343602(5/31/2027), 349005(10/31/2027), 374030(11/30/2027), 375434(12/31/2027), 377632(4/2/2028), 327194(8/31/2026), 375105(3/31/2028), 336074(3/21/2027); 81834B/20809160403324/326214(7/31/2027); 81834C/20809160466602/347551(8/12/2026), 372502(8/10/2026), 374615(10/21/2026), 376461(10/23/2026), 377088(11/6/2026), 373129(9/3/2026), 374033(9/3/2026)Quantity: 6,420Recall no. Z-3181-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide