PTW FREIBURG recalls Brand Name: VERIQA

Recall date · record updated Sep 23, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
PTW FREIBURG
Reason
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99720…

Products

  1. Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

    Codes/lots: Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;Quantity: 2Recall no. Z-3174-2026Class II