OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Medline Industries, LP recalls 2 products

Recall date · record updated Oct 14, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Medline Industries, LP
Reason
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide distribution - US Nationwide and the country of Canada.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99807…

Products (2)

  1. Rollator, Bariatric, Basic Steel, Burgundy, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWS

    Codes/lots: Lot Numbers: 05425090001 05425120001 05426010001 05426020001 05426030001 05426040001 05426050001Quantity: 1824 unitsRecall no. Z-0086-2027Class II
  2. Rollator, Bariatric, Basic Steel, Blue, mechanical walker with four legs and a metal frame intended for medical purposes to provide moderate weight support while walking, Product Number MDS86800XWBS

    Codes/lots: Lot Numbers: 05425080001 05426040001Quantity: 835 unitsRecall no. Z-0087-2027Class II

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API