Class IIOngoing
Beckman Coulter, Inc. recalls 2 products
Recall date · record updated Oct 21, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Beckman Coulter, Inc.
- Reason
- The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution including in the states and territory of CA, NE, FL, NY, MA, MD, SC, WI, TX, OH, VA, PA, IL, NH, KS, GA, NC, IA, PR, TN, OR, UT, NJ, MI, CO, NM, MS, LA, RI, OK, WA, MO, ME, IN, CT, AZ, AR, MT, MN, ID, DE, AL, WV, DC, KY, NV, WY, ND, HI, AK, SD, VT, GU and the countries of Canada, Mexico, Switzerland, Panama, Ecuador, Japan, Hungary, Italy, Belgium, China, Germany, France, United Kingdom of Great Britain and Northern Ireland, Russian Federation, Singapore, Romania, Argentina, Israel, Spain, Belarus, Guatemala, Colombia, Qatar, Czechia, Reunion, Australia, United Arab Emirates, Poland, Croatia, Brazil, Slovakia, New Zealand, Saudi Arabia, Thailand, Republic of Korea, Serbia, Austria, Macao, Ukraine, India, Netherlands, Taiwan, South Africa, Sint Maarten, Bangladesh, Hong Kong, Viet Nam, Bahrain, Uganda, Lebanon, Morocco, Guadeloupe, Greece, Curacao, Republic of Moldova, Malaysia, Ghana, Indonesia, Lithuania, Portugal, Kazakhstan, Georgia, Philippines, Bosnia and Herzegovina, Libya, Kuwait, Oman, Jordan, Uzbekistan.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99832…
Products (2)
DxU 840m Iris, REF: C76947; DxU 850m Iris, REF: C49513
Codes/lots: Product/UDI-DI: DxU 840m/15099590748654, DxU 850m/15099590748647. All Serial NumbersQuantity: 2,088Recall no. Z-0160-2027Class IIIQ200 SPRINT, REF: 700-3322,700-3322RFB, 700-3325, 700-3326, B48997; IQ200 SELECT, REF: 700-3340, 700-3345, 700-3347, B48998, C10684; IQ200 ELITE, REF: 700-3370, 700-3375, 700-3375RFB, 700-3376, B48999, C10683
Codes/lots: Product/UDI-DI: IQ200 SPRINT/10837461001256, IQ200 SELECT/10837461001300, IQ200 ELITE/10837461001355. All Serial NumbersRecall no. Z-0161-2027Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide