OVERLOOKED

Recalls › Medicines & devices

Class IOngoing

Ventec Life Systems, Inc. recalls 12 products

Recall date · record updated Oct 28, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Ventec Life Systems, Inc.
Reason
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Hazard
FDA Class I: a reasonable probability that use or exposure will cause serious adverse health consequences or death (FDA definition).
Who’s affected
Distribution US nationwide including Puerto Rico.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99835…

Products (12)

  1. VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007792, 00855573007006Quantity: 4,369 unitsRecall no. Z-0122-2027Class I
  2. VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007761, 00855573007204Quantity: 325 unitsRecall no. Z-0123-2027Class I
  3. VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007785, 00855573007433Quantity: 1,534 unitsRecall no. Z-0124-2027Class I
  4. VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007211Quantity: 606 unitsRecall no. Z-0125-2027Class I
  5. VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007747Quantity: 1,532 unitsRecall no. Z-0126-2027Class I
  6. VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00850018761130Quantity: 100 unitsRecall no. Z-0127-2027Class I
  7. VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007860Quantity: 4,368 unitsRecall no. Z-0128-2027Class I
  8. V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.

    Codes/lots: All Serial Numbers: PRT-01185-000 -UDI: 00855573007877; PRT-01185-001 - UDI: 00850018761161; PRT-01185-002 - UDI: 00850018761154Quantity: 39,568 unitsRecall no. Z-0129-2027Class I
  9. VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007730Quantity: 930 unitsRecall no. Z-0130-2027Class I
  10. VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00850018761116Quantity: 180 unitsRecall no. Z-0131-2027Class I
  11. V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007945Quantity: 30,120 unitsRecall no. Z-0132-2027Class I
  12. V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.

    Codes/lots: All Serial Numbers: UDI: 00855573007822Quantity: 25,112 unitsRecall no. Z-0133-2027Class I

Update history

  1. First listed on OVERLOOKED from the official record.

When the agency expands or changes this recall, we update this page and log the change here.

Proof: raw official record (JSON) · fetched from FDA API