Class IOngoing
Philips Ultrasound, LLC recalls Philips Lumify Diagnostic Ultrasound System
Recall date · record updated Oct 28, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Philips Ultrasound, LLC
- Reason
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- Hazard
- FDA Class I: a reasonable probability that use or exposure will cause serious adverse health consequences or death (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99878…
Products
Philips Lumify Diagnostic Ultrasound System
Codes/lots: UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.XQuantity: 13,206Recall no. Z-0152-2027Class I
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide