OVERLOOKED

Recalls › Medicines & devices

Class IIOngoing

Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI)

Recall date · record updated Aug 26, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)

What to do

Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.

Company
Fort Defiance Industries, LLC
Reason
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Hazard
FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
Who’s affected
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Where
Nationwide
Official notice
www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99371…

Products

  1. Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

    Codes/lots: Lot Code: All serial numbers/ GTIN: (01) 00862992000327Recall no. Z-2983-2026Class II