Class IIOngoing
Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI)
Recall date · record updated Aug 26, 2026 · Source: U.S. Food and Drug Administration (openFDA device enforcement)
Check your product against the names and codes above. The FDA enforcement record lists the recall as voluntary: firm initiated; it does not state a consumer remedy, so follow the firm’s notice or the FDA link.
- Company
- Fort Defiance Industries, LLC
- Reason
- The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
- Hazard
- FDA Class II: may cause temporary or medically reversible adverse health consequences (FDA definition).
- Who’s affected
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Where
- Nationwide
- Official notice
- www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99371…
Products
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Codes/lots: Lot Code: All serial numbers/ GTIN: (01) 00862992000327Recall no. Z-2983-2026Class II
Update history
- First listed on OVERLOOKED from the official record.
When the agency expands or changes this recall, we update this page and log the change here.
Proof: raw official record (JSON) · fetched from FDA API
More Medicines & devices recalls
Medicines & devicesClass I
Par Health USA, LLC recalls 2 products
Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
· FDA · Nationwide
Medicines & devicesClass II
Right Value Drug Stores LLC recalls Testosterone Cypionate
Lack of sterility assurance.
· FDA · Nationwide
Medicines & devicesClass II
STRIDES PHARMA INC recalls Cinacalcet Tablets
Failed dissolution specifications
· FDA · Nationwide